Aivia Software
Exocytosis Detection
The Exocytosis Detection recipe in Aivia detects exocytotic events in total internal reflection fluorescent (TIRF) microscopy images.
This recipe provides an option for tracking vesicles associated with exocytosis and is tunable for the detection of small, point exocytosis events (e.g., kiss-and-run) as well as large, diffuse events (e.g., full fusion).
On this page:
Parameters and Presets
Parameters
Recipe parameters for Exocytosis Detection and their descriptions are summarized in the table below.
Preset Group | Parameter Name | Min Value | Max Value | Description |
---|---|---|---|---|
Vesicle Detection | Vesicle Threshold (Vesicle Tracking only) | 1 | 255 | Adjusts the sensitivity for vesicle detection; a lower value will detect more vesicles |
Min Vesicle Size (Vesicle Tracking only) | 0 | 100 px2 or µm2 | Specifies the minimum size of a detected object to be included in the analysis results based on the areas of the detected objects | |
Event Detection
| Event Intensity Change | 0 | 255 (8-bit) 65,535 (16-bit) | Adjusts the sensitivity for event detection based on the size of the intensity change (in number of gray levels) at the event from baseline; a lower value will detect more events |
Transience | 0 | 100 | Specifies the range of duration for the events by computing a normalized ratio for the difference between the maximum and average intensities to the difference between the average and minimum intensities; a lower value will enable detection of events with longer durations | |
Bright Detection Frame (Enable Bright Detection only) | 0 | 255 (8-bit) 65,535 (16-bit) | Specifies the maximum number of frames that an event with extended intensity increase in its surrounding regions may last; this parameter is used only when Enable Bright Detection is selected | |
Duplicate Events Removal | Not applicable | Toggles the removal of detected events that occur in a given region in successive time frames | ||
Event Filter | Event Size Filter | 0 | 1,000 px2 or µm2 | Specifies the minimum size of a detected event to be included in the analysis results based on the area of the detected event |
Event Mean Intensity Filter | 0 | 255 (8-bit) 65,535 (16-bit) | Specifies the minimum threshold for mean intensity of a detected event region to be included in the analysis results | |
Event Total Intensity Filter (Enable Event Total Intensity Filter only) | 0 | 1 x 1018 | Specifies the minimum threshold for the total intensity of a detected event region to be included in the analysis results |
Presets
There are three preset groups in the recipe: Vesicle Detection, Event Detection and Event Filter. Each group has three pre-configured parameter groupings to help you get started on the analysis. The default preset values are as follows:
Vesicle Detection
Parameter Name | Dim | Medium | Bright |
---|---|---|---|
Vesicle Threshold | 15 | 25 | 40 |
Min Vesicle Size | 0.05 px2 or µm2 | 0.05 px2 or µm2 | 0.05 px2 or µm2 |
Event Detection
Parameter Name | Gross | Short | Long |
---|---|---|---|
Event Intensity Change | 8 (8-bit) | 8 (8-bit) | 8 (8-bit) |
Transience | 0 - 40 | 0 - 40 | 0 - 20 |
Bright Detection Frame | 0 | 2 | 5 |
Event Filter
Parameter Name | Sm-Bright | Medium | Large-Dim |
---|---|---|---|
Event Size Filter | 0.05 px2 or µm2 | 0.15 px2 or µm2 | 0.5 px2 or µm2 |
Event Mean Intensity Filter | 19 (8-bit) | 12 (8-bit) | 5 (8-bit) |
Event Total Intensity Filter | 15,000 | 45,000 | 50,000 |
Tutorial
Before beginning the tutorial, please download the Exocytosis Detection Demo image. For information on how to select presets or modify parameter values, please refer to the tutorial on how to use the Recipe Console.
Unzip the demo file and load the demo image, “ExocytosisDemo.tif,” into Aivia.
In the Recipe Console, click on the Recipe selection dropdown menu and select the Exocytosis Detection recipe.
Select the following presets for each preset group:
Vesicle Detection: Medium
Event Detection: Gross
Event Filter: Large-Dim
Click on the caret to the left of the Event Detection preset group to show the preset parameters.
Modify the parameter value as follows, leaving the other parameters intact:
Event Intensity Change (counts): 1,000
Click the Start button or press the
F4
key on your keyboard to begin applying the recipe to the image.
The detected particle tracks and outlines for the events will be overlaid on the image.
Results
Measurements
The Exocytosis Detection recipe generates intensity and position measurements for each detected event and, if Vesicle Tracking is enabled, for each detected particle in addition to morphology and advanced count measurements for the particles. You can add additional measurements to the analysis results with the Measurement Tool in Aivia and view measurement definitions on the Measurement Definitions page. The measurements generated by the recipe are given in the table below.
Object Set | Morphology | Intensity | Position | Advanced |
---|---|---|---|---|
Events | None |
|
| None |
Vesicle Tracks |
|
|
|
|
Image Credits
Takashi Tsuboi, University of Tokyo, Tokyo, Japan
Related Articles